HealthLab Flash-Master HFM-04
The HFM-04 HealthLab Flash-Master is a special version of the central controlling and storage module of the psycho-physiological monitoring system HealthLab.
In the context of measurements with HealthLab the HFM-04 Master controls by means of its Master-Module the functions of the HealthLab measuring satellites, which are connected via System-Bus. It receives the measurement data of the satellites, as well as the 3D acceleration values of a proband, for storage in its internal memory.
The HFM-04 captures the body surface temperature or the body core temperature of a proband, depending on the used temperature sensors ( NTC- or, as customer specific feature: double-temperature sensors ), as well as the probands ECG ( up to 3 channels ), the heart rate and the measuring data for pulse wave, temperature (skin surface) and electrical skin resistance of a finger sensor. The HFM-04 may store the captured data internally and transmits the data to a host device ( PC / smart-phone, Windows or 'Android' ), which is provided as control- and evaluation unit.
The HFM-04 is designed and optimised for permanent monitoring via Bluetooth connection to the control unit. The data rate of the Bluetooth LE interface is restricted to 8 kByte / sec.
The modular design of HealthLab Flash Master HFM-04 offers the opportunity to provide the device with an alternate satellite module, and thus with a different as herein described measurement technology of the monitoring system HealthLab, details on request.
Back to overview
TechniCAL DATA |
|
|---|---|
| Designation | HealthLab Flash-Master HFM-04, ECG (up to 3 channel), heart rate, electrical skin resistance, skin temperature, pulse wave, 3D acceleration, barometric air pressure, body core temperature |
| Power Supply / Consumption | internal: 2,0 … 3,0 V DC (2x Type AA) measuring without Bluetooth: 50 mA measuring with Bluetooth: 60 mA |
| Data Capturing | measuring ranges see Channel Index Datasheet Internal Sensors: |
| Digital Interfaces | Bluetooth 4.0 LE RS-485 HeallyBus using a four-core shielded cable 460/1840 kbps |
| Climatic Conditions | acc. to DIN EN 60204-1 ( 10-2014 ) |
| Ambient Temperature | operation: -20 … +55 °C transport / storage: -25 … +60 °C |
| Humidity / Altitude | 20 … 90 % RH ( without condensation ) up to 8.000 m |
| Dimensions | W / H / D : 65 x 127 x 23 mm weight : 120 g |
| Download PDF | |
| Ordering Information | HFM-04 Flash-Master: E1922 |
| Accessoires | TPS-02 Temperature Sensor: E1288 TcoreTM Temperature Sensor:AD0139 TcoreTM Adapter Cable: E1952 EGA-01 ECG Sensor Cable: E1938 ECG Strap: VV0069 VSS-03 Candy Cable (SATSAT connection cable): E1816 |